Synergy Machine Protocol  
 
Synergy Machine Protocol  
# Open Gen 5 software
+
#Open Gen 5 software
# Click on Existing protocol OR move to step 3
+
#Click on Existing protocol OR move to step 3
# Click on Create New Protocol  
+
#Click on Create New Protocol  
# Select Standard Protocol type
+
#Select Standard Protocol type
# Click on Procedure
+
#Click on Procedure
# Under Shake, select Orbital Shake for 3 seconds
+
#Under Shake, select Orbital Shake for 3 seconds
# Under Read, select Fluorescence from drop down menu
+
#Under Read, select Fluorescence from drop down menu
##       Enter Emission and Excitation wavelengths
+
##Enter Emission and Excitation wavelengths
###             If you are reading more than one type of fluorescence, click the appropriate number of types above the wavelength values.
+
###If you are reading more than one type of fluorescence, click the appropriate number of types above the wavelength values.
##       For RFP, select 585 nm (excitation) and 615 nm (emission) for optimal reading (with 100 gain)
+
##For RFP, select 585 nm (excitation) and 615 nm (emission) for optimal reading (with 100 gain)
##       For GFP, select 485 (excitation) and 515 (emission) for optimal reading (with 100 gain)
+
##For GFP, select 485 (excitation) and 515 (emission) for optimal reading (with 100 gain)
# Under Read, select Absorbance from the drop down menu and enter 590 nm  
+
##If GFP and RFP are combined in the same well, use a different method.
# Once those values have been submitted, drag tasks into proper order. First Shake, then read Fluorescence, then read Absorbance.
+
###Use the wavelengths of 460nm (excitation) and 490 nm(emission) for GFP, and 585 nm(excitation) and 615 nm (emission) for RFP, to minimize noise/interference. However, some of the GFP reading will be from bleeding over of RFP.       
##       If you are measuring GFP and RFP, measure GFP before RFP.
+
#Under Read, select Absorbance from the drop down menu and enter 590 nm  
# Click OK, and from the banner menu, select File, New Task, then Read Now (selecting the procedure you created)
+
#Once those values have been submitted, drag tasks into proper order. First Shake, then read Fluorescence, then read Absorbance.
# Place the well inside the machine, and Click OK when ready.
+
##If you are measuring GFP and RFP, measure GFP before RFP.
# After the machine has run, you can export the data to Excel using the Excel button on the banner of the window that pops up.
+
#Click OK, and from the banner menu, select File, New Task, then Read Now (selecting the procedure you created)
#       Note- if only a few wells give a value of OVRFLW, try lowering the gain from 100 to 95 or 90 (or lower if necessary). Enter the value on the page with the wavelengths for fluorescence. If all wells give a value of OVRFLW, this means the wavelengths are too close together. Try increasing the difference between the excitation and emission wavelengths.
+
#Place the well inside the machine, and Click OK when ready.
 +
#After the machine has run, you can export the data to Excel using the Excel button on the banner of the window that pops up.
 +
#Note- if only a few wells give a value of OVRFLW, try lowering the gain from 100 to 95 or 90 (or lower if necessary). Enter the value on the page with the wavelengths for fluorescence. If all wells give a value of OVRFLW, this means the wavelengths are too close together. Try increasing the difference between the excitation and emission wavelengths.
 
#[[File:Optimal GFP.PNG]]
 
#[[File:Optimal GFP.PNG]]
 
## 485 nm/515 nm (100 gain) are the optimal exitation/emission wavelengths for use with GFP.
 
## 485 nm/515 nm (100 gain) are the optimal exitation/emission wavelengths for use with GFP.
 
#[[File:CaptureRFP.PNG]]  
 
#[[File:CaptureRFP.PNG]]  
 
##585 nm/615 nm (100 gain) are the optimal excitation/emission wavelengths for use with RFP, as they resulted in the highest fluorescence readings.
 
##585 nm/615 nm (100 gain) are the optimal excitation/emission wavelengths for use with RFP, as they resulted in the highest fluorescence readings.

Revision as of 19:02, 7 June 2012

Synergy Machine Protocol

  1. Open Gen 5 software
  2. Click on Existing protocol OR move to step 3
  3. Click on Create New Protocol
  4. Select Standard Protocol type
  5. Click on Procedure
  6. Under Shake, select Orbital Shake for 3 seconds
  7. Under Read, select Fluorescence from drop down menu
    1. Enter Emission and Excitation wavelengths
      1. If you are reading more than one type of fluorescence, click the appropriate number of types above the wavelength values.
    2. For RFP, select 585 nm (excitation) and 615 nm (emission) for optimal reading (with 100 gain)
    3. For GFP, select 485 (excitation) and 515 (emission) for optimal reading (with 100 gain)
    4. If GFP and RFP are combined in the same well, use a different method.
      1. Use the wavelengths of 460nm (excitation) and 490 nm(emission) for GFP, and 585 nm(excitation) and 615 nm (emission) for RFP, to minimize noise/interference. However, some of the GFP reading will be from bleeding over of RFP.
  8. Under Read, select Absorbance from the drop down menu and enter 590 nm
  9. Once those values have been submitted, drag tasks into proper order. First Shake, then read Fluorescence, then read Absorbance.
    1. If you are measuring GFP and RFP, measure GFP before RFP.
  10. Click OK, and from the banner menu, select File, New Task, then Read Now (selecting the procedure you created)
  11. Place the well inside the machine, and Click OK when ready.
  12. After the machine has run, you can export the data to Excel using the Excel button on the banner of the window that pops up.
  13. Note- if only a few wells give a value of OVRFLW, try lowering the gain from 100 to 95 or 90 (or lower if necessary). Enter the value on the page with the wavelengths for fluorescence. If all wells give a value of OVRFLW, this means the wavelengths are too close together. Try increasing the difference between the excitation and emission wavelengths.
  14. Optimal GFP.PNG
    1. 485 nm/515 nm (100 gain) are the optimal exitation/emission wavelengths for use with GFP.
  15. CaptureRFP.PNG
    1. 585 nm/615 nm (100 gain) are the optimal excitation/emission wavelengths for use with RFP, as they resulted in the highest fluorescence readings.