Synergy Machine Protocol

From GcatWiki
Revision as of 21:04, 5 June 2012 by Cavrana (talk | contribs) (Created page with 'Synergy Machine Protocol 1. Open Gen 5 software 2. Click on Existing protocol OR move to step 3 3. Click on Create New Protocol 4. Select Standard Protocol type 5. Click on Pro…')
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

Synergy Machine Protocol 1. Open Gen 5 software 2. Click on Existing protocol OR move to step 3 3. Click on Create New Protocol 4. Select Standard Protocol type 5. Click on Procedure 6. Under Shake, select Orbital Shake for 3 seconds 7. Under Read, select Fluorescence from drop down menu a. Enter Emission and Excitation wavelengths i. If you are reading more than one type of fluorescence, click the appropriate number of types above the wavelength values. b. For RFP, select 585 nm (excitation) and 615 nm (emission) for optimal reading (with 100 gain) c. For GFP, select ___ (excitation) and ___ (emission) for optimal reading (with 100 gain) 8. Under Read, select Absorbance from the drop down menu and enter 590 nm 9. Once those values have been submitted, drag tasks into proper order. First Shake, then read Fluorescence, then read Absorbance. a. If you are measuring GFP and RFP, measure GFP before RFP. 10. Click OK, and from the banner menu, select File, New Task, then Read Now (selecting the procedure you created) 11. Place the well inside the machine, and Click OK when ready. 12. After the machine has run, you can export the data to Excel using the Excel button on the banner of the window that pops up.