Difference between revisions of "TAS2R38 PCR amplification"
From GcatWiki
Macampbell (talk | contribs) (Created page with '== PCR Amplify Your TAS2R38 Alleles == # When the DNA extraction cools, vortex the tubes for 30 seconds and then set up a new 500 µL microfuge tube by adding the following: {|…') |
Macampbell (talk | contribs) |
||
Line 1: | Line 1: | ||
− | |||
− | |||
# When the DNA extraction cools, vortex the tubes for 30 seconds and then set up a new 500 µL microfuge tube by adding the following: | # When the DNA extraction cools, vortex the tubes for 30 seconds and then set up a new 500 µL microfuge tube by adding the following: | ||
Revision as of 14:00, 7 August 2012
- When the DNA extraction cools, vortex the tubes for 30 seconds and then set up a new 500 µL microfuge tube by adding the following:
Reagent | Volume |
---|---|
extracted DNA | 10.5 µL |
2X Clear GoTaq Mix | 12.5 µL |
Forward Primer | 1.0 µL |
Reverse Primer | 1.0 µL |
Final Volume | 25.0 µL |
2. Run the thermocycler program called TAS2R38 with the heated lid enabled. PCR conditions:
- 5 minutes at 95˚C.
- 1 minute at 95˚C.
- 1 minute at 65˚C.
- 1 minute at 72˚C.
- repeat steps 2 - 5 29 more times.
- hold at room temperature.
- END
3. When the PCR is completed, store DNA frozen until you are ready to clean up the DNA in order to have your two alleles sequenced.