Difference between revisions of "Synergy Machine Protocol for Bio113"
From GcatWiki
Macampbell (talk | contribs) (Created page with " == Synergy Machine Protocol == # All Wells should contain 200 µL of liquid. Be sure to include wells containing broth with no cells for negative control. # We use the wavel...") |
Macampbell (talk | contribs) |
||
Line 2: | Line 2: | ||
== Synergy Machine Protocol == | == Synergy Machine Protocol == | ||
− | # All | + | # All wells should contain exactly 200 µL of liquid. Be sure to include wells containing broth with no cells for negative control. |
− | # We use the wavelengths of | + | # We use the wavelengths of 460 nm (excitation) and 490 nm(emission) for GFP, and 585 nm(excitation) and 615 nm (emission) for RFP, to minimize noise/interference. However, some of the GFP reading will be from bleeding over of RFP. |
# If you are measuring GFP and RFP, measure GFP before RFP. | # If you are measuring GFP and RFP, measure GFP before RFP. | ||
+ | # Choose the protocol called '''Bio113 Synthetic Biology Research''' | ||
# Place the 96-well plate inside the machine, and be sure to put well A1 near the marker for A1. | # Place the 96-well plate inside the machine, and be sure to put well A1 near the marker for A1. | ||
# After the machine has run, you can export the data to Excel using the Excel button on the banner of the window that pops up. | # After the machine has run, you can export the data to Excel using the Excel button on the banner of the window that pops up. |
Revision as of 20:57, 18 July 2017
Synergy Machine Protocol
- All wells should contain exactly 200 µL of liquid. Be sure to include wells containing broth with no cells for negative control.
- We use the wavelengths of 460 nm (excitation) and 490 nm(emission) for GFP, and 585 nm(excitation) and 615 nm (emission) for RFP, to minimize noise/interference. However, some of the GFP reading will be from bleeding over of RFP.
- If you are measuring GFP and RFP, measure GFP before RFP.
- Choose the protocol called Bio113 Synthetic Biology Research
- Place the 96-well plate inside the machine, and be sure to put well A1 near the marker for A1.
- After the machine has run, you can export the data to Excel using the Excel button on the banner of the window that pops up.
Note- if only a few wells give a value of OVRFLW, try lowering the gain from 100 to 95 or 90 (or lower if necessary). Enter the value on the page with the wavelengths for fluorescence.