Synergy Machine Protocol for Bio113

From GcatWiki
Revision as of 20:55, 18 July 2017 by Macampbell (talk | contribs) (Created page with " == Synergy Machine Protocol == # All Wells should contain 200 µL of liquid. Be sure to include wells containing broth with no cells for negative control. # We use the wavel...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

Synergy Machine Protocol

  1. All Wells should contain 200 µL of liquid. Be sure to include wells containing broth with no cells for negative control.
  2. We use the wavelengths of 460nm (excitation) and 490 nm(emission) for GFP, and 585 nm(excitation) and 615 nm (emission) for RFP, to minimize noise/interference. However, some of the GFP reading will be from bleeding over of RFP.
  3. If you are measuring GFP and RFP, measure GFP before RFP.
  4. Place the 96-well plate inside the machine, and be sure to put well A1 near the marker for A1.
  5. After the machine has run, you can export the data to Excel using the Excel button on the banner of the window that pops up.



Note- if only a few wells give a value of OVRFLW, try lowering the gain from 100 to 95 or 90 (or lower if necessary). Enter the value on the page with the wavelengths for fluorescence.